NewTom GO (Legacy), Wall-Mount Site Spec

General Specifications

ELECTRICAL115V OR 230V. Unit comes configured for 230V. Included power cable has NEMA 5-15-P
Dedicated 20A circuit
DATA2 CAT6 lines run from pan location to acquisition PC location
Acquisition PC requires a gigabit Ethernet connection to customer’s network
EXPOSURE SWITCHCoiled cord switch included. Switch can be relocated anywhere by running a network line. Unit is compatible with 2-wire doorbell-style buttons.

Dimensions and Measurements

BACKING IN WALL
IF BACKING IS NEEDED IT SHOULD EXTEND FROM ABOUT
39″ TO 60″ OFF THE GROUND, AT LEAST 18″ WIDE.
DEVICE DIMENSIONS
MIN RECOMMENDED ROOM DIMENSIONS 46″D X 55.5″W
WALL PLATE WITH STUDS
-Plate 17.75″W and 19.75″T
-Mounting holes are 16″ on-center
-Oval eyelets are aprx 1″ wide
-Plate has 4 threaded rods for mounting the column