NewTom GiANO HR Site Spec

General Specifications

ELECTRICAL115V OR 230V. Unit comes configured for 230V. Included power cable has NEMA 5-15-P
Dedicated 20A circuit
DATA2 CAT6 lines run from pan location to acquisition PC location
Acquisition PC requires a gigabit Ethernet connection to customer’s network
EXPOSURE SWITCHCoiled cord switch included. Switch can be relocated anywhere by running a network line. Unit is compatible with 2-wire doorbell-style buttons.
BACKING IN WALL
Center of backing is 46″ off the ground
DEVICE DIMENSIONS
Ceph RIGHT
WALL BRACKETDEVICE DIMENSIONS
Ceph LEFT

Mounting Options

Floor PlatesWall Brackets
  1. Easy Access Base Plate
  2. Standard Base Plate
  3. N/A
  4. Standard Wall Mount Bracket
  5. Single-Stud Wall Mount Bracket
  6. 45-degree Wall Mount Bracket

Note: The bracket or half-column that supports the X-ray unit must be firmly fixed to the wall. Select a fastening system suitable for the wall capable of withstanding a 220 Kg tear force applied at each anchoring point.